SPECIFICATIONS(TEC207/215)
PARAMETER |
MODEL |
UNIT |
|||
TEC207L |
TEC207 |
TEC215L |
TEC215 |
||
Supports Sensors |
NTC PT1000/PT100 |
|
|||
Measurable Temperature Range |
-200~800 |
°C |
|||
Temperature Sensitivity |
0.001 |
0.0001 |
0.001 |
0.0001 |
°C |
Temperature Measurement Drift Caused by Ambient Temperature |
0.0001 |
°C/°C |
|||
Optimal Temperature Control Stability (related to the overall system) |
±0.01 |
±0.001 |
±0.01 |
±0.001 |
°C |
Communication Method |
485 Serial Port (supports Modbus and ASCII communication protocols) |
|
|||
Power Supply Voltage Range (DC) |
7~24 |
V |
|||
Output Polarity |
Bipolar、Unipolar |
|
|||
Number of Channels |
2 |
|
|||
Maximum Output Voltage |
Settable |
|
|||
Output Current Range (SSR: Solid State Relay) |
±7A each channel |
±15A each channel 0-±80A/SSR |
A |
||
Output Current Limit |
±10A |
±20A |
A |
||
Operating Ambient Temperature |
-55~60 |
°C |
|||
Operating Ambient Humidity |
0~98 |
%RH |
|||
Thermal Dissipation Requirements |
No Additional Thermal Dissipation Needed Within Rated Operating Range |
|
|||
Circuit Board Overheat Protection |
Yes |
|
|||
Power Loss Memory |
Yes |
|
|||
PID Parameters |
User Adjustable |
|
|||
Dimension |
94.3*79.5*20.5 |
mm |
|||
Weight |
≈240 |
g |
Dimensional Drawing
。
Interfaces
Pin Name |
Pin Type |
Pin Definition (High Level: 3.3V, Low Level: 0V) |
|
① |
ENABLE |
Input |
Thermostat Enable Terminal. High Level: The thermostat can output normally; Low Level: The thermostat is not allowed to output. |
STATE |
Output |
Thermostat Overtemperature Alarm Signal Output. High Level: (1) The thermostat's own temperature is greater than its overtemperature threshold; (2) The temperature of sensor 1 or 2 exceeds their respective high and low threshold ranges. Low Level: No overtemperature anomaly. |
|
GND |
Input |
Power Supply Input Negative (Low Current) |
|
② |
VIN |
Input |
Power Supply Input Positive (Low Current). |
485A |
Output |
RS485 Signal Line A. Data Bits: 8 bits, Stop Bits: 1 bit, Parity: None, Baud Rate: 9600. |
|
485B |
Output |
RS485 Signal Line B. Data Bits: 8 bits, Stop Bits: 1 bit, Parity: None, Baud Rate: 9600. |
|
GND |
Input |
Power Supply Input Negative (Low Current). |
|
③ |
TXD |
Output |
Serial Port Transmitter, TTL Level, used for connecting to PC control software. Data Bits: 8, Stop Bit: 1, Parity: None, Baud Rate: 38400. |
RXD |
Input |
Serial Port Receiver, TTL Level, used for connecting to PC control software. Data Bits: 8, Stop Bit: 1, Parity: None, Baud Rate: 38400. |
|
GND |
Input |
Power Supply Input Negative (Low Current). |
|
④ |
TEC1+ |
Output |
The positive terminal of the temperature control current output is typically connected to the positive terminal of the Thermoelectric Cooler (TEC). |
TEC1- |
Output |
The negative terminal of the temperature control current output is usually connected to the negative terminal of the Thermoelectric Cooler (TEC). |
|
⑤ |
VIN |
Input |
Power Input Positive Pole, with an input voltage range of 7 to 24V. |
GND |
Input |
Power Supply Input Negative (High Current). |
|
⑥ |
TEC2+ |
Output |
The positive terminal of the temperature control current output is typically connected to the positive terminal of the Thermoelectric Cooler (TEC). |
TEC2- |
Output |
The negative terminal of the temperature control current output is usually connected to the negative terminal of the Thermoelectric Cooler (TEC). |
|
⑦ |
1k |
Input |
Platinum Resistance Thermometer (Pt1000) Interface (Please contact Technical Support before purchase). |
COM |
Input |
Common Interface for Platinum Resistance Thermometer (Pt1000) and NTC Thermistor. |
|
10k |
Input |
Thermistor (NTC) Interface. |
|
⑧ |
1k |
Input |
Platinum Resistance Thermometer (Pt1000) Interface (Please contact Technical Support before purchase). |
COM |
Input |
Common Interface for Platinum Resistance Thermometer (Pt1000) and NTC Thermistor. |
|
10k |
Input |
Thermistor (NTC) Interface. |
|
⑨ |
PWM1 |
Output |
Solid State Relay Input Terminal 1 |
PWM2 |
Output |
Solid State Relay Input Terminal 2 |
|
GND |
Input |
Power Supply Input Negative (Low Current) |
|
PWM3 |
Output |
Solid State Relay Input Terminal 3 |
|
PWM4 |
Output |
Solid State Relay Input Terminal 4 |
|
GND |
Input |
Power Supply Input Negative (Low Current) |
|
⑩ |
VCC |
Output |
Buzzer Alarm Positive. Outputs high level when there is an error in temperature control. |
GND |
Output |
Buzzer Alarm Negative. |